RBS3202 Test with CMU300 and RNC/Iub SIM
1CM56_E 8 Rohde & Schwarz
Transmitter measurements
R&S CMU300:
Start the function group 3G UMTS Node B / WCDMA 1900 / FDD /
Non-Signalling Mode on the R&S CMU300.
Reset the activated WCDMA function group (hardkey RESET).
Configure the instrument for measurement (softkey Connect. Control)
as follows:
Reset the instrument to external reference frequency (tab Sync. /
Reference Frequency: Ext. (at REF IN); 10 MHz).
Configure the RF paths (tab AF / RF: RF 1 is RF input and RF
output; external attenuation must be taken into account as
appropriate).
Set the DL scrambling code to be used on the RNC/Iub simulator
(tab Node B / DL Channel Settings / Scrambling Code / Primary: 0;
Secondary: 0 ;these are default values and normally must not be
changed).
Switch to the Measurement menus (softkey Connect. Control) and
enter the following settings in the Power Measurement menu (WCDMA
FDD POWER):